Free Shipping on orders over US$39.99 mail:[email protected]

Main Parts and Materials of High Vacuum Flapper Valve

SaveSavedRemoved 0
Deal Score0
Deal Score0

Valve body, bonnet and stem: aluminum alloy, carbon steel or stainless steel

Seals: NBR, Viton, bellows (shaft seal).

Product features

  1. Standardized, modular design, easy to replace and maintain;
  2. Easy to clean design;
  3. The solenoid valve adopts energy-saving design and small volume.

Main performance indexes

  1. Scope of application

Pressure difference on valve plate when opening: ≤ 1.2 × 105Pa in any direction

Leakage rate of valve body and valve seat: ≤ 1.3 × 10-7pa · L · S-1

Recommended number of cycles before first maintenance: 200000

Baking temperature of valve body: ≤ 150 ℃

  1. Installation position

Manual drive: any – pneumatic drive: any

Electromagnetic drive: any (bellows seal); seal face to vacuum (fluorine rubber ring seal)

  1. 3.power supply

Pneumatic drive: AC 220V 50Hz, 6W or DC 24V, 3W (special specification can be customized)

Electromagnetic drive: AC 220V or 230V 50Hz, starting power 800W, holding power 8W

Compressed air (pneumatic only): 0.4 ~ 0.7MPa

Valve opening and closing time – pneumatic drive: ≤ 1s

Electromagnetic drive: opening ≤ 0.1s; closing ≤ 1s (fluororubber ring seal), closing ≤ 0.5s (bellows seal)

  1. Valve position indication

Manual drive: with opening and closing position indication

Pneumatic drive: with opening and closing position indicator switch (magnetic switch)

Electromagnetic drive: with on position indicator

Labideal
Logo
Register New Account
Compare items
  • Total (0)
Compare