Labideal GPR-7500 AIS ATEX PPM H2S Analyzer ATEX II
Features
This analyzer features an advanced trace PPM hydrogen sulfide sensor capable of precisely measuring hydrogen sulfide concentrations ranging from 1 – 200 PPM over a 24 month operating sensor life with no periodic maintenance required, menu driven controls, two field selectable alarm set points.
- Calibration – span gas 50 – 100 PPM H2S balance air
- 0.2 – 1 scfh flow sensitivity
- Compensation – barometric pressure and temperature
- Controls – menu driven range selection, calibration, alarms and system functions
- +/-1% full scale linearity
- Response time – 90% of final reading in 35 sec
- Less than 0.5% full scale sensitivity
- Stainless steel wetted parts
- Sensor does not freeze at -20 deg C
- Sample system includes sample flow meter, sample/span control valve
- Pressure regulate to 5 – 30 psig inlet and atmospheric vent
- Sensor – model OSV-72-HH-LM, 24 months expected life at 25 deg C, 1 atmosphere and 12 months warranty time
- Alarm type – two user configurable, 3 A at 100 VAC relay, time delay, bypass
- 12 months warranty
Specifications
- Brand Name: Labideal
- UNSPSC: 41113100
- Item: Hydrogen Sulfide Analyzer
- Series: GPR-7500 AIS ATEX
- Type: Intrinsically Safe
- Accuracy: Less than 2%
- Detection Range: 0 – 20 PPM, 0 – 50 PPM and 0 – 100 PPM
- Alarm Type: Two User Configurable
- Signal Output: 4 – 20 mA
- Process Connection: 1/8 Inch Compression Tube
- Display: 0.1 PPM Resolution 2-3/4 x 1-3/8 Inch Graphical LCD
- Pressure Rating: 5 – 30 psig
- Operating Temperature: -10 to 45 deg C
- Power Source: 12 – 24 VDC or 100/230 VAC
- Enclosure Type: NEMA 4X, Explosion Proof
- Material: Aluminum/Fiberglass
- Dimensions: 21 x 9 x 6 Inch
- Process Application: Trace PPM Hydrogen Sulfide in flowing gas streams
- Standard/Approvals: ATEX 94/9/EC II 2 G, Ex [ib] ib IIB T4 Gb, CE Certified, ISO Certified