Labideal GPR-2800 AIS-LD ATEX Oxygen Analyzer Sample Sys
Features
- Calibration – 3 month interval using air or certified span gas of O2 approximating approx 80% of full scale range balance N2
- Barometric pressure compensation
- Controls – water resistant keypad, menu driven range selection, calibration and system functions
- Flow sensitivity – none, 0.4 – 0.9 Nlpm recommended
- +/-1% full scale linearity
- Pressure regulate 0.3 – 2 barg to deliver 0.4 – 0.9 Nlpm flow to transmitter inlet and atmospheric vent
- Response time – 90% of final reading in 10 sec
- Less than 0.5% full scale sensitivity
- Stainless steel wetted parts
- Sensor – model GPR-11-32 for gases containing >0.5% (CO2), 32 months expected life at 25 deg C, 1 atmosphere and 12 months warranty time
- 12 months warranty
Specifications
- Brand Name: Labideal
- UNSPSC: 41113110
- Item: Oxygen Analyzer
- Series: GPR-2800 AIS-LD ATEX
- Type: Intrinsically Safe
- Accuracy: Less than 2%
- Detection Range: 0 – 1%, 0 – 5%, 0 – 10% and 0 – 25%
- Alarm Type: Two User Configurable
- Signal Output: 4 – 20 mA, Modbus RTU Communication
- Process Connection: 1/8 Inch Compression Tube
- Display: 0.01 PPM Resolution 70 x 35 mm Graphical LCD
- Pressure Rating: 0.3 – 2 barg
- Operating Temperature: -10 to 45 deg C
- Power Source: 110 – 220 VAC
- Enclosure Type: NEMA 3R, Outdoor
- Includes: Liquid Drain System
- Process Application: Oxygen Analysis in inert, hydrocarbon, helium, hydrogen, mixed and acid (CO2) gas streams
- Standard/Approvals: ATEX Directive 94/9/EC II 2 G Ex d [ib] ib IIB T4