Labideal GPR-28 ATEX Explosion Proof Oxygen Analyzer ATEX
Features
The proprietary oxygen sensor capable of precisely measuring oxygen concentrations ranging from 0.001 – 100% over a 32 months operating sensor life with no periodic maintenance, integral temperature control enhances stability, a proven electronic platform and a stainless steel bypass sample system that enables the analyzer to come online in minutes.
- Calibration – certified gas of O2 balance N2 or ambient air
- Flow sensitivity – none, between 1 – 5 scfh, 2 scfh recommended
- Temperature compensation
- Controls – explosion proof actuators for range selection, zero and span calibration adjustments
- >0.995% over all ranges linearity
- Response time – 90% of final reading in <10 sec
- Less than 0.5% full scale sensitivity
- Stainless steel wetted parts
- Pressure regulate to 5-30 psig inlet, 150 psig maximum and atmospheric vent
- Sensor – model GPR-11-32, 32 months expected life at 25 deg C, 1 atmosphere and 12 months warranty time
- Alarm type – two adjustable form C non-latching relay contacts, sensor and power failure
- Area classification – enclosure Class 1, Division 1, Groups B, C and D
- 12 months warranty
Specifications
- Brand Name: Labideal
- UNSPSC: 41113110
- Item: Oxygen Analyzer
- Series: GPR-28 ATEX
- Type: Explosion Proof
- Accuracy: Less than 2%
- Detection Range: 0 – 1%, 0 – 5%, 0 – 10% and 0 – 25%
- Alarm Type: Two Adjustable Form C Non-Latching Relay Contacts
- Signal Output: 0 – 1 V and 4 – 20 mA Isolated
- Process Connection: 1/8 Inch x 1/4 Inch Compression Tube
- Display: 0.001% Resolution 3-1/2 Digits Bright Red LED
- Pressure Rating: 150 psig
- Operating Temperature: -10 to 45 deg C
- Power Source: 100/120 VAC or 220/240 VAC
- Enclosure Type: Wall Mount
- Material: Painted Aluminum
- Dimensions: 16 x 18 x 11 Inch
- Process Application: Oxygen Analysis in hydrocarbon, hydrogen, mixed and acid (CO2) gas streams
- Standard/Approvals: ATEX Directive 94/9/EC Ex II 2 G Ex d IIB, H2 T5 or T6, CE Certified