Labideal GPR-2500 SB Oxygen Transmitter
Features
This transmitter features an advanced galvanic oxygen sensor capable of measuring oxygen concentrations ranging from 0.05 – 100% over a 60 month operating life with no periodic maintenance required, menu driven controls and 2-wire loop power.
- Calibration – air or certified span gas of O2 balance N2 approximating 80% of range above analysis range recommended
- Flow sensitivity – none, between 0.5 – 5 scfh, 1 – 2 scfh recommended
- Recovery time – 60 sec in air to <10 PPM in <1 hr on N2 purge
- Barometric pressure and temperature compensation
- Controls – water resistant keypad, menu driven range selection, calibration and system functions
- +/-1% full scale linearity
- Response time – 90% of final reading in 10 sec
- Less than 0.5% full scale sensitivity
- Stainless steel wetted parts
- Auto-ranging or manually lock on single range
- Pressure regulate to 5 – 30 psig to deliver 1 – 2 scfh flow to transmitter inlet and atmospheric vent
- Sensor – model GPR-12-333 for inert gases and XLT-12-333 for gases containing >0.5% CO2, 24 months expected life at 25 deg C, 1 atmosphere and 12 months warranty time
- 12 months analyzer warranty
Specifications
- Brand Name: Labideal
- UNSPSC: 41111924
- Item: Oxygen Transmitter
- Type: Intrinsically Safe
- Series: GPR-2500 SB
- Detection Range: 0 – 10 PPM, 0 – 100 PPM, 0 – 1000 PPM, 0 – 1% and 0 – 25%
- Accuracy: Less than 1%
- Signal Output: 4 – 20 mA
- Process Connection: 1/8 Inch Compression Tube
- Power Source: 12 – 24 VDC, 2-Wire Looped
- Pressure Rating: 5 – 30 psig
- Display: 0.01 PPM Resolution 2-3/4 x 1-3/8 Inch Graphical LCD
- Enclosure Type: NEMA 4X
- Temperature Rating: -10 to 45 deg C (GPR), -20 to 45 deg C (XLT)
- Dimensions: 23 x 10 x 8 cm
- Material: Painted Aluminum
- Process Application: Oxygen Analysis in inert, hydrocarbon, hydrogen, mixed and acid (CO2) gas streams
- Standard/Approvals: CE Certified