Labideal GPR-2500 AIS-LD ATEX Oxygen Analyzer/Sample Sys
Features
This analyzer features an advanced galvanic oxygen sensor capable of measuring oxygen concentrations ranging from 0.05 – 100% over a 36 month operating life, menu driven controls, two field selectable alarm set points, and, a stainless steel sample system.
- Calibration – air or certified span gas of O2 balance N2 approximating 80% of range above analysis range recommended
- Flow sensitivity – none, between 0.5 – 5 scfh, 1 – 2 scfh recommended
- Barometric pressure and temperature compensation
- Controls – water resistant keypad, menu driven range selection, calibration and system functions
- +/-1% full scale linearity
- Response time – 90% of final reading in 10 sec
- Less than 0.5% full scale sensitivity
- Stainless steel wetted parts
- Pressure regulate 5 – 30 psig to deliver 1-2 scfh flow to transmitter inlet and atmospheric vent
- Sensor – model XLT-12-333 for gases containing >0.5% CO2, 24 months expected life at 25 deg C, 1 atmosphere and 12 months warranty time
- Alarm type – two user configurable, magnetic coil relays rated 3 A at 100 VAC, field programmable, time delays, bypass for calibration and system failure
- Auto-ranging or manually lock on single range
- Sample system – sample flow meter
- Area classification – meets the recognized intrinsic safety standards for use in Class 1, Division 1, Group C and D hazardous areas
- 12 months warranty
Specifications
- Brand Name: Labideal
- UNSPSC: 41113110
- Item: Oxygen Analyzer
- Series: GPR-2500 AIS-LD ATEX
- Type: Intrinsically Safe
- Accuracy: Less than 2%
- Detection Range: 0 – 10 PPM, 0 – 100 PPM, 0 – 1000 PPM, 0 – 1% and 0 – 25%
- Alarm Type: Two User Configurable
- Signal Output: 4 – 20 mA
- Process Connection: 1/8 Inch Compression Tube
- Display: 0.01 PPM Resolution 2-3/4 x 1-3/8 Inch Graphical LCD
- Pressure Rating: 5 – 30 psig
- Operating Temperature: -20 to 45 deg C
- Power Source: 12 – 24 VDC or 100/230 VAC
- Enclosure Type: NEMA 4X, Explosion Proof
- Material: Aluminum/Fiberglass
- Dimensions: 16 x 9 x 6 Inch
- Includes: Liquid Drain System
- Process Application: Oxygen Analysis in inert, hydrocarbon, helium, hydrogen, mixed and acid (CO2) gas streams
- Standard/Approvals: ATEX Directive 94/9/EC II 2 G Ex d [ib] ib IIB T4, CE Certified